Background. Gatekeeper (GK) has shown to be safe and effective in patients with fecal incontinence (FI). We aimed to understand its mechanism of action by comparing pre- and post-implant change in the external anal sphincter (EAS) contractility. Methods. Study of EAS contractility was conducted in 16 FI females (median age = 69 years) before and after implant of 6 GK prostheses. Muscle tension (Tm), expressed in millinewtons per centimeter squared, mN(cm2)−1, was calculated using the equation Tm = P(ri)(tm)−1, where P is the average maximum squeeze pressure and ri and tm the inner radius and thickness of the EAS, respectively. The effect of a predefined set of covariates on Tm was tested by restricted maximum likelihood models. Results. Compared with baseline, despite unchanged tm (2.7 [2.5-2.8] vs 2.5 [2.2-2.8] mm; P =.31 mm), a significant increase in P (median = 45.8 [26.5-75.8] vs 60.4 [43.1-88.1] mm Hg; P =.017), and ri (12.4 [11.5-13.4] vs 18.7 [17.3-19.6] mm; P <.001) resulted in an increase in Tm (233.2 [123.8-303.2] vs 490.8 [286.9-562.4] mN(cm2)−1; P <.001) at 12 months after GK implant. Twelve-month follow-up improvements were also observed on Cleveland Clinic FI score (8-point median decrease; P =.0001), St Marks FI score (10-point median decrease; P <.0001), and American Medical Systems score (39-point median decrease; P <.0001). Restricted maximum likelihood models showed that years of onset of FI was negatively associated with change in Tm (P =.048). Conclusions. GK-related EAS compression positively influences muscle contractility by increasing ri, with consequent increase in Tm (length-tension relationship). Further studies are needed to confirm the long-term effectiveness of GK.

Grossi, U., De Simone, V., Parello, A., Litta, F., Donisi, L., Di Tanna, G. L., Goglia, M., Ratto, C., Gatekeeper Improves Voluntary Contractility in Patients With Fecal Incontinence, <<SURGICAL INNOVATION>>, 2019; 2019 (26/3): 321-327. [doi:10.1177/1553350618818924] [http://hdl.handle.net/10807/181213]

Gatekeeper Improves Voluntary Contractility in Patients With Fecal Incontinence

Parello, Angelo;Litta, Francesco;Donisi, Lorenza;Ratto, Carlo
2018

Abstract

Background. Gatekeeper (GK) has shown to be safe and effective in patients with fecal incontinence (FI). We aimed to understand its mechanism of action by comparing pre- and post-implant change in the external anal sphincter (EAS) contractility. Methods. Study of EAS contractility was conducted in 16 FI females (median age = 69 years) before and after implant of 6 GK prostheses. Muscle tension (Tm), expressed in millinewtons per centimeter squared, mN(cm2)−1, was calculated using the equation Tm = P(ri)(tm)−1, where P is the average maximum squeeze pressure and ri and tm the inner radius and thickness of the EAS, respectively. The effect of a predefined set of covariates on Tm was tested by restricted maximum likelihood models. Results. Compared with baseline, despite unchanged tm (2.7 [2.5-2.8] vs 2.5 [2.2-2.8] mm; P =.31 mm), a significant increase in P (median = 45.8 [26.5-75.8] vs 60.4 [43.1-88.1] mm Hg; P =.017), and ri (12.4 [11.5-13.4] vs 18.7 [17.3-19.6] mm; P <.001) resulted in an increase in Tm (233.2 [123.8-303.2] vs 490.8 [286.9-562.4] mN(cm2)−1; P <.001) at 12 months after GK implant. Twelve-month follow-up improvements were also observed on Cleveland Clinic FI score (8-point median decrease; P =.0001), St Marks FI score (10-point median decrease; P <.0001), and American Medical Systems score (39-point median decrease; P <.0001). Restricted maximum likelihood models showed that years of onset of FI was negatively associated with change in Tm (P =.048). Conclusions. GK-related EAS compression positively influences muscle contractility by increasing ri, with consequent increase in Tm (length-tension relationship). Further studies are needed to confirm the long-term effectiveness of GK.
2018
Inglese
Grossi, U., De Simone, V., Parello, A., Litta, F., Donisi, L., Di Tanna, G. L., Goglia, M., Ratto, C., Gatekeeper Improves Voluntary Contractility in Patients With Fecal Incontinence, <<SURGICAL INNOVATION>>, 2019; 2019 (26/3): 321-327. [doi:10.1177/1553350618818924] [http://hdl.handle.net/10807/181213]
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/10807/181213
Citazioni
  • ???jsp.display-item.citation.pmc??? 3
  • Scopus 15
  • ???jsp.display-item.citation.isi??? 12
social impact