In uncooperative patients, electrical compound action potential (ECAP) thresholds are reliable in predicting T-levels, but are not in determining the C-level profile. The present study aims to assess if the C-level profile can be predicted by a new objective procedure (C-NRT) which uses the amplitude growth function (AGF) and is based on the assumption that equal ECAP amplitudes elicit equal loudness percepts. This is a correlational study conducted in five tertiary care referral hospitals with 21 post-lingually deaf adult cochlear implant users. Two maps were created: a behavioral, bitonal balanced (BB) map and an objective map, in which T-levels were the same as in the BB map, and C-levels were obtained with C-NRT. C-NRT consisted of performing the AGF of nine electrodes, and of setting the current level eliciting a 100 μV ECAP amplitude as C-level in the map. AutoNRT was also measured. Main outcome measures were correlation between behavioral C-profile level, objective C-profile level, behavioral T-profile level and objective T-profile (AutoNRT) level; disyllabic word recognition scores in quiet and in noise conditions (SNR = + 10 and 0) with both maps. A strong correlation was found between behavioral and C-NRT-derived C-levels (mean per electrode correlation: R = 0.862, p < 0.001). C-NRT could predict behavioral C-levels with a greater accuracy than AutoNRT. Word recognition was significantly better with BB maps only in the quiet condition (p = 0.002). C-NRT is more accurate than AutoNRT in predicting the C-level profile in adult cochlear implant users. This finding encourages future application in uncooperative patients, especially in very young children.

Scorpecci, A., D’Elia, A., Malerba, P., Cantore, I., Consolino, P., Trabalzini, F., Paludetti, G., Quaranta, N. A. A., Maps created using a new objective procedure (C-NRT) correlate with behavioral, loudness-balanced maps: a study in adult cochlear implant users, <<EUROPEAN ARCHIVES OF OTO-RHINO-LARYNGOLOGY>>, 2016; 273 (12): 4167-4173. [doi:10.1007/s00405-016-4115-1] [http://hdl.handle.net/10807/95518]

Maps created using a new objective procedure (C-NRT) correlate with behavioral, loudness-balanced maps: a study in adult cochlear implant users

Scorpecci, Alessandro
;
Cantore, Italo;Paludetti, Gaetano
Penultimo
;
Quaranta, Nicola Antonio Adolfo
Ultimo
2016

Abstract

In uncooperative patients, electrical compound action potential (ECAP) thresholds are reliable in predicting T-levels, but are not in determining the C-level profile. The present study aims to assess if the C-level profile can be predicted by a new objective procedure (C-NRT) which uses the amplitude growth function (AGF) and is based on the assumption that equal ECAP amplitudes elicit equal loudness percepts. This is a correlational study conducted in five tertiary care referral hospitals with 21 post-lingually deaf adult cochlear implant users. Two maps were created: a behavioral, bitonal balanced (BB) map and an objective map, in which T-levels were the same as in the BB map, and C-levels were obtained with C-NRT. C-NRT consisted of performing the AGF of nine electrodes, and of setting the current level eliciting a 100 μV ECAP amplitude as C-level in the map. AutoNRT was also measured. Main outcome measures were correlation between behavioral C-profile level, objective C-profile level, behavioral T-profile level and objective T-profile (AutoNRT) level; disyllabic word recognition scores in quiet and in noise conditions (SNR = + 10 and 0) with both maps. A strong correlation was found between behavioral and C-NRT-derived C-levels (mean per electrode correlation: R = 0.862, p < 0.001). C-NRT could predict behavioral C-levels with a greater accuracy than AutoNRT. Word recognition was significantly better with BB maps only in the quiet condition (p = 0.002). C-NRT is more accurate than AutoNRT in predicting the C-level profile in adult cochlear implant users. This finding encourages future application in uncooperative patients, especially in very young children.
2016
Inglese
Scorpecci, A., D’Elia, A., Malerba, P., Cantore, I., Consolino, P., Trabalzini, F., Paludetti, G., Quaranta, N. A. A., Maps created using a new objective procedure (C-NRT) correlate with behavioral, loudness-balanced maps: a study in adult cochlear implant users, <<EUROPEAN ARCHIVES OF OTO-RHINO-LARYNGOLOGY>>, 2016; 273 (12): 4167-4173. [doi:10.1007/s00405-016-4115-1] [http://hdl.handle.net/10807/95518]
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/10807/95518
Citazioni
  • ???jsp.display-item.citation.pmc??? 3
  • Scopus 7
  • ???jsp.display-item.citation.isi??? 7
social impact