BACKGROUND: Continuous glucose monitoring using subcutaneous sensors has been validated in adults and children with diabetes, and was found to be useful in the management of glucose control. We aimed to assess feasibility and reliability of a new continuous glucose monitoring system (CGMS) in a population of preterm neonates using a Clarke error grid (CEG) specifically modified for preterm infants. METHODS: Preterm infants were recruited within 24 h from delivery. A subcutaneous sensor connected to a CGMS was inserted and maintained for 6 days. Data collected from CGMS were compared with data obtained using a glucometer. Management of the infants followed standard protocols and was not influenced by CGMS readings. RESULTS: Twenty patients (9 males) were included. Median (range) gestational age was 32 weeks (27-36) and median (range) birth weight was 1350 g (860-3360). Average CGMS recording time was 137 h, for a total of 449 paired glucose levels. CEG and modified CEG criteria for clinical significance were met. CONCLUSION: CGMS is a safe and clinically adequate method to estimate glucose levels in preterm infants. As the glucose level can be evaluated in real time, this CGMS could be useful to reduce the number of heel sticks, to observe glycaemic trends and to promptly detect episodes of both hypo- and hyper-glycaemia.

Tiberi, E., Cota, F., Barone, G., Perri, A., Romano, V., Iannotta, R., Romagnoli, C., Zecca, E., Continuous glucose monitoring in preterm infants: evaluation by a modified Clarke error grid., <<THE ITALIAN JOURNAL OF PEDIATRICS>>, 2016; (Mar 9;42): 29-29. [doi:10.1186/s13052-016-0236-9] [http://hdl.handle.net/10807/95077]

Continuous glucose monitoring in preterm infants: evaluation by a modified Clarke error grid.

Tiberi, Eloisa
Primo
;
Cota, Francesco
Secondo
;
Barone, Giovanni;Perri, Alessandro;Romano, Valerio;Iannotta, Rossella;Romagnoli, Costantino
Penultimo
;
Zecca, Enrico
Ultimo
2016

Abstract

BACKGROUND: Continuous glucose monitoring using subcutaneous sensors has been validated in adults and children with diabetes, and was found to be useful in the management of glucose control. We aimed to assess feasibility and reliability of a new continuous glucose monitoring system (CGMS) in a population of preterm neonates using a Clarke error grid (CEG) specifically modified for preterm infants. METHODS: Preterm infants were recruited within 24 h from delivery. A subcutaneous sensor connected to a CGMS was inserted and maintained for 6 days. Data collected from CGMS were compared with data obtained using a glucometer. Management of the infants followed standard protocols and was not influenced by CGMS readings. RESULTS: Twenty patients (9 males) were included. Median (range) gestational age was 32 weeks (27-36) and median (range) birth weight was 1350 g (860-3360). Average CGMS recording time was 137 h, for a total of 449 paired glucose levels. CEG and modified CEG criteria for clinical significance were met. CONCLUSION: CGMS is a safe and clinically adequate method to estimate glucose levels in preterm infants. As the glucose level can be evaluated in real time, this CGMS could be useful to reduce the number of heel sticks, to observe glycaemic trends and to promptly detect episodes of both hypo- and hyper-glycaemia.
2016
Inglese
Tiberi, E., Cota, F., Barone, G., Perri, A., Romano, V., Iannotta, R., Romagnoli, C., Zecca, E., Continuous glucose monitoring in preterm infants: evaluation by a modified Clarke error grid., <<THE ITALIAN JOURNAL OF PEDIATRICS>>, 2016; (Mar 9;42): 29-29. [doi:10.1186/s13052-016-0236-9] [http://hdl.handle.net/10807/95077]
File in questo prodotto:
File Dimensione Formato  
13052_2016_Article_236.pdf

accesso aperto

Tipologia file ?: Versione Editoriale (PDF)
Licenza: Creative commons
Dimensione 971.17 kB
Formato Adobe PDF
971.17 kB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/10807/95077
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 20
  • ???jsp.display-item.citation.isi??? 20
social impact