BackgroundMonitoring urinary albumin is a useful method in clinical practice for the management of diabetic nephropathy, chronic kidney disease, and hypertension. Currently there are neither standardized methods nor reference material for the determination of urinary albumin; for this reason it is useful to compare different assays used in clinical laboratory. ObjectivesThe aim of this study is to verify analytical performance of an immunoturbidimetric assay on Roche Cobas 8000 platform and to compare urinary albumin results with those obtained by immunonephelometry on Siemens Dade Behring BN II Nephelometer. ResultsThe method comparison showed a good linear relationship, confirmed by Passing-Bablok and Bland-Altman plots. The turbidimetric assay meets the requirements of accuracy and precision for the practice of medical diagnostics and clinical use. ConclusionsThe present study can contribute to the methods standardization and harmonization of urinary albumin assay.

Molinario, R., Pocino, K., Daloiso, P. D., Giannace, A., Spirito, G., Zuppi, C., Antenucci, M., Urinary Albumin Detection: Comparison of Two Different Methods, <<JOURNAL OF CLINICAL LABORATORY ANALYSIS>>, 2016; 30 (6): 888-891. [doi:10.1002/jcla.21952] [http://hdl.handle.net/10807/94575]

Urinary Albumin Detection: Comparison of Two Different Methods

Molinario, Rossana
Primo
;
Pocino, Krizia
Secondo
;
Giannace, Angela;Zuppi, Cecilia
Penultimo
;
Antenucci, Mirca
Ultimo
2016

Abstract

BackgroundMonitoring urinary albumin is a useful method in clinical practice for the management of diabetic nephropathy, chronic kidney disease, and hypertension. Currently there are neither standardized methods nor reference material for the determination of urinary albumin; for this reason it is useful to compare different assays used in clinical laboratory. ObjectivesThe aim of this study is to verify analytical performance of an immunoturbidimetric assay on Roche Cobas 8000 platform and to compare urinary albumin results with those obtained by immunonephelometry on Siemens Dade Behring BN II Nephelometer. ResultsThe method comparison showed a good linear relationship, confirmed by Passing-Bablok and Bland-Altman plots. The turbidimetric assay meets the requirements of accuracy and precision for the practice of medical diagnostics and clinical use. ConclusionsThe present study can contribute to the methods standardization and harmonization of urinary albumin assay.
2016
Inglese
Molinario, R., Pocino, K., Daloiso, P. D., Giannace, A., Spirito, G., Zuppi, C., Antenucci, M., Urinary Albumin Detection: Comparison of Two Different Methods, <<JOURNAL OF CLINICAL LABORATORY ANALYSIS>>, 2016; 30 (6): 888-891. [doi:10.1002/jcla.21952] [http://hdl.handle.net/10807/94575]
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/10807/94575
Citazioni
  • ???jsp.display-item.citation.pmc??? 3
  • Scopus 8
  • ???jsp.display-item.citation.isi??? 7
social impact