Electronic noses (e-noses), artificial sensor systems generally consisting of chemical sensor arrays for the detection of volatile compound profiles, have potential applications in respiratory medicine. We assessed within-day and between-day repeatability of an e-nose made from 32 sensors in patients with stable chronic obstructive pulmonary disease (COPD). We also compared between-day repeatability of an e-nose, fraction of exhaled nitric oxide (FNO) and pulmonary function testing. Within-day and between-day repeatability for the e-nose was assessed in two breath samples collected 30 min and seven days apart, respectively. Repeatability was expressed as an intraclass correlation coefficient (ICC). All sensors had ICC above 0.5, a value that is considered acceptable for repeatability. Regarding within-day repeatability, ICC ranged from 0.75 to 0.84 (mean = 0.80 ± 0.004). Sensors 6 and 19 were the most reproducible sensors (both, ICC = 0.84). Regarding between-day repeatability, ICC ranged from 0.57 to 0.76 (mean = 0.68 ± 0.01). Sensor 19 was the most reproducible sensor (ICC = 0.76). Within-day e-nose repeatability was greater than between-day repeatability (P < 0.0001). Between-day repeatability of FNO (ICC = 0.91) and spirometry (ICC range = 0.94-0.98) was greater than that of e-nose (mean ICC = 0.68). In patients with stable COPD, the e-nose used in this study has acceptable within-day and between-day repeatability which varies between different sensors.

Bofan, M., Mores, N., Dabrowska, M., Valente, S., Schmid, M., Trove', A., Conforto, S., Zini, G., Cattani Franchi, P., Mondino, C., Pagliari, G., D'Alessio, T., Montuschi, P., Mautone, A., Fuso, L., WHITIN-DAY AND BETWEEN-DAY REPEATIBILITY OF MEASURERMENTS WITH AN ELECTRONIC NOSE IN PATIENTS WITH COPD, <<JOURNAL OF BREATH RESEARCH>>, 2013; 7 (Marzo): N/A-N/A. [doi:10.1088/1752-7155/7/1/017103] [http://hdl.handle.net/10807/43171]

WHITIN-DAY AND BETWEEN-DAY REPEATIBILITY OF MEASURERMENTS WITH AN ELECTRONIC NOSE IN PATIENTS WITH COPD

Mores, Nadia;Cattani Franchi, Paola;Montuschi, Paolo;
2013

Abstract

Electronic noses (e-noses), artificial sensor systems generally consisting of chemical sensor arrays for the detection of volatile compound profiles, have potential applications in respiratory medicine. We assessed within-day and between-day repeatability of an e-nose made from 32 sensors in patients with stable chronic obstructive pulmonary disease (COPD). We also compared between-day repeatability of an e-nose, fraction of exhaled nitric oxide (FNO) and pulmonary function testing. Within-day and between-day repeatability for the e-nose was assessed in two breath samples collected 30 min and seven days apart, respectively. Repeatability was expressed as an intraclass correlation coefficient (ICC). All sensors had ICC above 0.5, a value that is considered acceptable for repeatability. Regarding within-day repeatability, ICC ranged from 0.75 to 0.84 (mean = 0.80 ± 0.004). Sensors 6 and 19 were the most reproducible sensors (both, ICC = 0.84). Regarding between-day repeatability, ICC ranged from 0.57 to 0.76 (mean = 0.68 ± 0.01). Sensor 19 was the most reproducible sensor (ICC = 0.76). Within-day e-nose repeatability was greater than between-day repeatability (P < 0.0001). Between-day repeatability of FNO (ICC = 0.91) and spirometry (ICC range = 0.94-0.98) was greater than that of e-nose (mean ICC = 0.68). In patients with stable COPD, the e-nose used in this study has acceptable within-day and between-day repeatability which varies between different sensors.
2013
Inglese
Bofan, M., Mores, N., Dabrowska, M., Valente, S., Schmid, M., Trove', A., Conforto, S., Zini, G., Cattani Franchi, P., Mondino, C., Pagliari, G., D'Alessio, T., Montuschi, P., Mautone, A., Fuso, L., WHITIN-DAY AND BETWEEN-DAY REPEATIBILITY OF MEASURERMENTS WITH AN ELECTRONIC NOSE IN PATIENTS WITH COPD, <<JOURNAL OF BREATH RESEARCH>>, 2013; 7 (Marzo): N/A-N/A. [doi:10.1088/1752-7155/7/1/017103] [http://hdl.handle.net/10807/43171]
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/10807/43171
Citazioni
  • ???jsp.display-item.citation.pmc??? 19
  • Scopus 67
  • ???jsp.display-item.citation.isi??? 63
social impact