Background: The aim of this study was to investigate the accuracy in fibrosis staging of a novel shear wave elastography (SWE) device (S-Shearwave Imaging by Samsung) and a previously validated 2D-SWE by Supersonic Imagine (SSI) in patients with biopsy proven metabolic dysfunction-associated steatotic liver disease (MASLD). Methods: This prospective study included 75 consecutive patients with MASLD who underwent liver biopsy for suspected MASH. All patients underwent S-Shearwave Imaging by Samsung and 2D-SWE with SSI on the same day of liver biopsy. Fibrosis was histologically assessed using the METAVIR classification system. Agreement between the equipment was assessed with the Pearson coefficient. A receiver operator characteristic curve (ROC) analysis with the Youden index was used to establish thresholds for fibrosis staging. Results: A good correlation was found between S-Shearwave Imaging by Samsung and 2D-SWE with SSI (Pearson’s R = 0.68; p < 0.01). At multivariate regression analysis, S-Shearwave Imaging was associated with advanced fibrosis (≥F3) independently from age, diabetes and platelets (OR 2.94, CI 1.69–5.11, p < 0.01). The fibrosis diagnostic accuracy of both S-Shearwave Imaging and 2D-SWE was good to optimal with AUROCs of 0.81 and 0.70 for significant fibrosis (≥F2), 0.94 and 0.91 for severe fibrosis (≥F3), respectively. The accuracy of S-Shearwave is not significantly different from Fibroscan and Agile3+ (DeLong test p value 0.16 and 0.15, respectively) while is slightly better than 2D-SWE, FIB4 and NFS (DeLong test p value < 0.05). For S-Shearwave Imaging by Samsung, the best cut-off values for diagnosing fibrosis ≥F2, ≥F3 were, respectively, 7.9 kPa (Sens 74.4%, Spec 87.5%) and 8.1 kPa (Sens 95.6%, Spec 78.8%). For 2D-SWE by SSI, the best cut-off values for diagnosing fibrosis ≥F2, ≥F3 were, respectively, 7.2 kPa (Sens 55.8%, Spec 84.4%) and 7.6 kPa (Sens 82.6%, Spec 84.6%). Conclusion: S-Shearwave Imaging is a useful and reliable non-invasive technique for staging liver fibrosis in patients with MASLD. Its diagnostic accuracy is non-inferior to other shear wave elastography techniques (TE and 2D-SWE by SSI).

Liguori, A., Esposto, G., Ainora, M. E., Mignini, I., Borriello, R., Galasso, L., Paratore, M., Giustiniani, M. C., Riccardi, L., Garcovich, M., Gasbarrini, A., Miele, L., Zocco, M. A., Liver Elastography for Liver Fibrosis Stratification: A Comparison of Three Techniques in a Biopsy-Controlled MASLD Cohort, <<BIOMEDICINES>>, N/A; 13 (1): N/A-N/A. [doi:10.3390/biomedicines13010138] [https://hdl.handle.net/10807/330767]

Liver Elastography for Liver Fibrosis Stratification: A Comparison of Three Techniques in a Biopsy-Controlled MASLD Cohort

Esposto, Giorgio;Ainora, Maria Elena;Mignini, Irene;Borriello, Raffaele;Galasso, Linda;Paratore, Mattia;Giustiniani, Maria Cristina;Riccardi, Laura;Garcovich, Matteo;Gasbarrini, Antonio;Miele, Luca;Zocco, Maria Assunta
2025

Abstract

Background: The aim of this study was to investigate the accuracy in fibrosis staging of a novel shear wave elastography (SWE) device (S-Shearwave Imaging by Samsung) and a previously validated 2D-SWE by Supersonic Imagine (SSI) in patients with biopsy proven metabolic dysfunction-associated steatotic liver disease (MASLD). Methods: This prospective study included 75 consecutive patients with MASLD who underwent liver biopsy for suspected MASH. All patients underwent S-Shearwave Imaging by Samsung and 2D-SWE with SSI on the same day of liver biopsy. Fibrosis was histologically assessed using the METAVIR classification system. Agreement between the equipment was assessed with the Pearson coefficient. A receiver operator characteristic curve (ROC) analysis with the Youden index was used to establish thresholds for fibrosis staging. Results: A good correlation was found between S-Shearwave Imaging by Samsung and 2D-SWE with SSI (Pearson’s R = 0.68; p < 0.01). At multivariate regression analysis, S-Shearwave Imaging was associated with advanced fibrosis (≥F3) independently from age, diabetes and platelets (OR 2.94, CI 1.69–5.11, p < 0.01). The fibrosis diagnostic accuracy of both S-Shearwave Imaging and 2D-SWE was good to optimal with AUROCs of 0.81 and 0.70 for significant fibrosis (≥F2), 0.94 and 0.91 for severe fibrosis (≥F3), respectively. The accuracy of S-Shearwave is not significantly different from Fibroscan and Agile3+ (DeLong test p value 0.16 and 0.15, respectively) while is slightly better than 2D-SWE, FIB4 and NFS (DeLong test p value < 0.05). For S-Shearwave Imaging by Samsung, the best cut-off values for diagnosing fibrosis ≥F2, ≥F3 were, respectively, 7.9 kPa (Sens 74.4%, Spec 87.5%) and 8.1 kPa (Sens 95.6%, Spec 78.8%). For 2D-SWE by SSI, the best cut-off values for diagnosing fibrosis ≥F2, ≥F3 were, respectively, 7.2 kPa (Sens 55.8%, Spec 84.4%) and 7.6 kPa (Sens 82.6%, Spec 84.6%). Conclusion: S-Shearwave Imaging is a useful and reliable non-invasive technique for staging liver fibrosis in patients with MASLD. Its diagnostic accuracy is non-inferior to other shear wave elastography techniques (TE and 2D-SWE by SSI).
2025
Inglese
Liguori, A., Esposto, G., Ainora, M. E., Mignini, I., Borriello, R., Galasso, L., Paratore, M., Giustiniani, M. C., Riccardi, L., Garcovich, M., Gasbarrini, A., Miele, L., Zocco, M. A., Liver Elastography for Liver Fibrosis Stratification: A Comparison of Three Techniques in a Biopsy-Controlled MASLD Cohort, <<BIOMEDICINES>>, N/A; 13 (1): N/A-N/A. [doi:10.3390/biomedicines13010138] [https://hdl.handle.net/10807/330767]
File in questo prodotto:
File Dimensione Formato  
biomedicines-13-00138.pdf

accesso aperto

Tipologia file ?: Versione Editoriale (PDF)
Licenza: Creative commons
Dimensione 1.04 MB
Formato Adobe PDF
1.04 MB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/10807/330767
Citazioni
  • ???jsp.display-item.citation.pmc??? 2
  • Scopus 2
  • ???jsp.display-item.citation.isi??? 2
social impact