Aim: This ex vivo study aims to compare the accuracy of two electronic apex locators, Wirele-X (Forum Engineering Technologies Ltd., Israel), tested in two modalities, alone and connected to the display screen and Dentaport ZX (J. Morita, Tokyo, Japan) in determining the working length. Methodology: 15 single-rooted teeth were selected for this study. The actual microscopic working length was measured using a size 10 K-file that was advanced until the tip of the file could be visualized just within the apical foramen under a stereomicroscope. Then each tooth was placed into the alginate to simulate the clinical conditions. Wirele-X and Dentaport ZX were used according manufacturer's instructions. Three measurements were performed for each tooth and each apex locator and modality and differences between the electronic and actual working lengths were calculated. Positive values indicated measurements that extruded beyond the apical foramen, while negative values indicated measurements that were short of the apical foramen. Means and standard deviation were calculated and the statistical analysis was performed using One-way ANOVA and Tukey tests (P<0.05). Results: The difference between electronic and actual working length was 0.05±0.34 mm for Wirele-X alone, 0.003±0.37 mm for Wirele-X connected to the display screen and 0.08±0.35 mm for Dentaport ZX. No statistical differences were found among Wirele-X alone, Wirele-X connected to the display screen and Dentaport ZX (p>0.05). Conclusions: Wirele-X in both modalities and Dentaport ZX showed a high accuracy in determining working length and were accurate to within ±0.5 mm, without any statistical differences among them.

Castagnola, R., Diana, R., Colangeli, M., Panzetta, C., Marigo, L., Grande, N. M., Plotino, G., Effectiveness of a new electronic apex locator in two modalities in detecting the working length: An ex vivo study, <<GIORNALE ITALIANO DI ENDODONZIA>>, 2021; 35 (1): 10-16. [doi:10.32067/GIE.2021.35.01.21] [https://hdl.handle.net/10807/242174]

Effectiveness of a new electronic apex locator in two modalities in detecting the working length: An ex vivo study

Castagnola, Raffaella;Diana, Rosalba;Marigo, Luca;Grande, Nicola Maria;Plotino, Gianluca
2021

Abstract

Aim: This ex vivo study aims to compare the accuracy of two electronic apex locators, Wirele-X (Forum Engineering Technologies Ltd., Israel), tested in two modalities, alone and connected to the display screen and Dentaport ZX (J. Morita, Tokyo, Japan) in determining the working length. Methodology: 15 single-rooted teeth were selected for this study. The actual microscopic working length was measured using a size 10 K-file that was advanced until the tip of the file could be visualized just within the apical foramen under a stereomicroscope. Then each tooth was placed into the alginate to simulate the clinical conditions. Wirele-X and Dentaport ZX were used according manufacturer's instructions. Three measurements were performed for each tooth and each apex locator and modality and differences between the electronic and actual working lengths were calculated. Positive values indicated measurements that extruded beyond the apical foramen, while negative values indicated measurements that were short of the apical foramen. Means and standard deviation were calculated and the statistical analysis was performed using One-way ANOVA and Tukey tests (P<0.05). Results: The difference between electronic and actual working length was 0.05±0.34 mm for Wirele-X alone, 0.003±0.37 mm for Wirele-X connected to the display screen and 0.08±0.35 mm for Dentaport ZX. No statistical differences were found among Wirele-X alone, Wirele-X connected to the display screen and Dentaport ZX (p>0.05). Conclusions: Wirele-X in both modalities and Dentaport ZX showed a high accuracy in determining working length and were accurate to within ±0.5 mm, without any statistical differences among them.
2021
Inglese
Castagnola, R., Diana, R., Colangeli, M., Panzetta, C., Marigo, L., Grande, N. M., Plotino, G., Effectiveness of a new electronic apex locator in two modalities in detecting the working length: An ex vivo study, <<GIORNALE ITALIANO DI ENDODONZIA>>, 2021; 35 (1): 10-16. [doi:10.32067/GIE.2021.35.01.21] [https://hdl.handle.net/10807/242174]
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/10807/242174
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 0
  • ???jsp.display-item.citation.isi??? 0
social impact