Aims: In heart failure (HF), implantable hemodynamic monitoring devices have been shown to optimize therapy, anticipating clinical decompensation and preventing hospitalization. Direct left-sided hemodynamic sensors offer theoretical benefits beyond pulmonary artery pressure (PAP) monitoring systems. We evaluated the safety, usability, and performance of a novel left atrial pressure (LAP) monitoring system in HF patients. Methods and results: The VECTOR-HF study(NCT03775161) was a first-in-human, prospective, multicenter, single-arm, clinical trial enrolling 30 patients with HF. The device consisted of an interatrial positioned leadless sensor, able to transmit LAP data wirelessly. After three months, a right heart catheterization (RHC) was performed to correlate mean pulmonary capillary wedge pressure (PCWP) with simultaneous mean LAP obtained from the device. Remote LAP measurements were then used to guide patient management. The miniaturized device was successfully implanted in all 30 patients, without acute Major Adverse Cardiac and Neurological Events (MACNE). At 3 months, freedom from short-term MACNE was 97%. Agreement between sensor-calculated LAP and PCWP was consistent, with a mean difference of -0.22±4.92mmHg, the correlation coefficient and the Lin's Concordance Correlation Coefficient values were equal to 0.79 (P<0.0001) and 0.776 (95%CI=0.582-0.886), respectively. Preliminary experience with VLAP-based HF management was associated with significant improvements in NYHA functional class (32% of patients reached NYHA II class at 6 months, P<0.005; 60% of patients at 12 months, P<0.005) and 6-minute walk-test distance (from 244.59±119.59m at baseline to 311.78±129.88m after 6 months, P<0.05, and 343.95±146.15m after 12 months, P<0.05). Conclusion: The V-LAP™ monitoring system proved to be generally safe and provided a good correlation with invasive PCWP. Initial evidence also suggests possible improvement in HF clinical symptoms. This article is protected by copyright. All rights reserved.

D'Amario, D., Meerkin, D., Restivo, A., Ince, H., Sievert, H., Wiese, A., Schaefer, U., Trani, C., Bayes-Genis, A., Leyva, F., Whinnett, Z. I., Di Mario, C., Jonas, M., Manhal, H., Amat-Santos, I. J., Del Trigo, M., Gal, T. B., Ben Avraham, B., Hasin, T., Feickert, S., D'Ancona, G., Altisent, O. A., Koren, O., Caspi, O., Abraham, W. T., Crea, F., Anker, S. D., Kornowski, R., Perl, L., Safety, usability, and performance of a wireless left atrial pressure monitoring system in patients with heart failure: the VECTOR-HF trial, <<EUROPEAN JOURNAL OF HEART FAILURE>>, 2023; (24 apr): N/A-N/A. [doi:10.1002/ejhf.2869] [https://hdl.handle.net/10807/234451]

Safety, usability, and performance of a wireless left atrial pressure monitoring system in patients with heart failure: the VECTOR-HF trial

D'Amario, Domenico;Restivo, Attilio;Trani, Carlo;Crea, Filippo;
2023

Abstract

Aims: In heart failure (HF), implantable hemodynamic monitoring devices have been shown to optimize therapy, anticipating clinical decompensation and preventing hospitalization. Direct left-sided hemodynamic sensors offer theoretical benefits beyond pulmonary artery pressure (PAP) monitoring systems. We evaluated the safety, usability, and performance of a novel left atrial pressure (LAP) monitoring system in HF patients. Methods and results: The VECTOR-HF study(NCT03775161) was a first-in-human, prospective, multicenter, single-arm, clinical trial enrolling 30 patients with HF. The device consisted of an interatrial positioned leadless sensor, able to transmit LAP data wirelessly. After three months, a right heart catheterization (RHC) was performed to correlate mean pulmonary capillary wedge pressure (PCWP) with simultaneous mean LAP obtained from the device. Remote LAP measurements were then used to guide patient management. The miniaturized device was successfully implanted in all 30 patients, without acute Major Adverse Cardiac and Neurological Events (MACNE). At 3 months, freedom from short-term MACNE was 97%. Agreement between sensor-calculated LAP and PCWP was consistent, with a mean difference of -0.22±4.92mmHg, the correlation coefficient and the Lin's Concordance Correlation Coefficient values were equal to 0.79 (P<0.0001) and 0.776 (95%CI=0.582-0.886), respectively. Preliminary experience with VLAP-based HF management was associated with significant improvements in NYHA functional class (32% of patients reached NYHA II class at 6 months, P<0.005; 60% of patients at 12 months, P<0.005) and 6-minute walk-test distance (from 244.59±119.59m at baseline to 311.78±129.88m after 6 months, P<0.05, and 343.95±146.15m after 12 months, P<0.05). Conclusion: The V-LAP™ monitoring system proved to be generally safe and provided a good correlation with invasive PCWP. Initial evidence also suggests possible improvement in HF clinical symptoms. This article is protected by copyright. All rights reserved.
2023
Inglese
D'Amario, D., Meerkin, D., Restivo, A., Ince, H., Sievert, H., Wiese, A., Schaefer, U., Trani, C., Bayes-Genis, A., Leyva, F., Whinnett, Z. I., Di Mario, C., Jonas, M., Manhal, H., Amat-Santos, I. J., Del Trigo, M., Gal, T. B., Ben Avraham, B., Hasin, T., Feickert, S., D'Ancona, G., Altisent, O. A., Koren, O., Caspi, O., Abraham, W. T., Crea, F., Anker, S. D., Kornowski, R., Perl, L., Safety, usability, and performance of a wireless left atrial pressure monitoring system in patients with heart failure: the VECTOR-HF trial, <<EUROPEAN JOURNAL OF HEART FAILURE>>, 2023; (24 apr): N/A-N/A. [doi:10.1002/ejhf.2869] [https://hdl.handle.net/10807/234451]
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/10807/234451
Citazioni
  • ???jsp.display-item.citation.pmc??? 0
  • Scopus ND
  • ???jsp.display-item.citation.isi??? ND
social impact