OBJECTIVES: To compare the accuracy of a new transcutaneous bilirubinometer, BiliMed (Medick SA, Paris, France) with BiliCheck (Respironics, Marietta, GA, USA), a widely available instrument, and with total serum bilirubin measurement. DESIGN: A prospective double-blind study comparing the two devices was carried out. 686 healthy newborns needing measurement of their bilirubin were enrolled over a 4-month period. Serum and transcutaneous bilirubin measurements were taken with both devices within 15 minutes. The order of use of the instruments was randomised. SETTING: Well-baby nursery ward in a university hospital, tertiary referral centre. RESULTS: The linear regression analysis showed a better correlation between BiliCheck and serum bilirubin (r = 0.75) than between BiliMed and serum bilirubin (r = 0.45). BiliCheck variability (+/-2 SD of the mean bias from serum bilirubin) was within -87.2 to 63.3 micromol/l, while BiliMed variability was within -97.5 to 121.4 micromol/l. The receiver operating characteristic analysis (for serum bilirubin levels >205.2 micromol/l or >239.4 micromol/l) showed significantly higher areas under the curve for BiliCheck than those for BiliMed (p<0.001). CONCLUSIONS: Despite the potential practical advantages of BiliMed, its reduced diagnostic accuracy in comparison with BiliCheck does not justify its use in clinical practice.

De Luca, D., Zecca, E., Corsello, M., Tiberi, E., Semeraro, C., Romagnoli, C., Attempt to improve transcutateous bilirubinometry: a double blinded study Medick BiliMed versus Respironic BiliCheck, <<ARCHIVES OF DISEASE IN CHILDHOOD. FETAL AND NEONATAL EDITION>>, 2007; (Nov 15): 135-139 [http://hdl.handle.net/10807/22787]

Attempt to improve transcutateous bilirubinometry: a double blinded study Medick BiliMed versus Respironic BiliCheck

De Luca, Daniele;Zecca, Enrico;Romagnoli, Costantino
2007

Abstract

OBJECTIVES: To compare the accuracy of a new transcutaneous bilirubinometer, BiliMed (Medick SA, Paris, France) with BiliCheck (Respironics, Marietta, GA, USA), a widely available instrument, and with total serum bilirubin measurement. DESIGN: A prospective double-blind study comparing the two devices was carried out. 686 healthy newborns needing measurement of their bilirubin were enrolled over a 4-month period. Serum and transcutaneous bilirubin measurements were taken with both devices within 15 minutes. The order of use of the instruments was randomised. SETTING: Well-baby nursery ward in a university hospital, tertiary referral centre. RESULTS: The linear regression analysis showed a better correlation between BiliCheck and serum bilirubin (r = 0.75) than between BiliMed and serum bilirubin (r = 0.45). BiliCheck variability (+/-2 SD of the mean bias from serum bilirubin) was within -87.2 to 63.3 micromol/l, while BiliMed variability was within -97.5 to 121.4 micromol/l. The receiver operating characteristic analysis (for serum bilirubin levels >205.2 micromol/l or >239.4 micromol/l) showed significantly higher areas under the curve for BiliCheck than those for BiliMed (p<0.001). CONCLUSIONS: Despite the potential practical advantages of BiliMed, its reduced diagnostic accuracy in comparison with BiliCheck does not justify its use in clinical practice.
2007
Inglese
De Luca, D., Zecca, E., Corsello, M., Tiberi, E., Semeraro, C., Romagnoli, C., Attempt to improve transcutateous bilirubinometry: a double blinded study Medick BiliMed versus Respironic BiliCheck, <<ARCHIVES OF DISEASE IN CHILDHOOD. FETAL AND NEONATAL EDITION>>, 2007; (Nov 15): 135-139 [http://hdl.handle.net/10807/22787]
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/10807/22787
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus ND
  • ???jsp.display-item.citation.isi??? ND
social impact