OBJECTIVES: To compare the accuracy of BiliCheck™ (Respironics, Marietta, GA) and Konica-Minolta Air Shield JM-103 (Drager Medical Inc, Telford, PA) to evaluate total serum bilirubin (TSB). METHODS: Prospective blinded study comparing two diagnostic devices in 630 neonates requiring TSB measurement. RESULTS: Linear regression analysis showed a good correlation between BiliCheck™ and TSB (r=0.8212) as well as between JM-103 and TSB (r=0.8686). BiliCheck shows a tendency to underestimate TSB. The mean difference in TSB-TcB was -1.4 mg/dL for BC (-4.7/+1.8 mg/dL) and 0.3 mg/dL for JM-103 (-2.6/+3.2mg/dL). ROC analysis for TSB≥ 12 mg/dL showed area under the curve for BiliCheck™ significantly lower than those for JM-103 (p<0.0001). JM-103 resulted less time expensive than BiliCheck. CONCLUSIONS: In spite of similar diagnostic accuracy JM-103 could be preferred for some practical advantages, but its suitability in performing universal screening for severe hyperbilirubinemia deserves further investigations.

Romagnoli, C., Zecca, E., Catenazzi, P., Barone, G., Zuppa, A. A., Transcutaneous bilirubin measurement: Comparison of Respironics BiliCheck and JM-103 in a normal newborn population, <<CLINICAL BIOCHEMISTRY>>, 2012; 45 (9): 659-662. [doi:10.1016/j.clinbiochem.2012.03.014] [http://hdl.handle.net/10807/21374]

Transcutaneous bilirubin measurement: Comparison of Respironics BiliCheck and JM-103 in a normal newborn population

Romagnoli, Costantino;Zecca, Enrico;Catenazzi, Piero;Barone, Giovanni;Zuppa, Antonio Alberto
2012

Abstract

OBJECTIVES: To compare the accuracy of BiliCheck™ (Respironics, Marietta, GA) and Konica-Minolta Air Shield JM-103 (Drager Medical Inc, Telford, PA) to evaluate total serum bilirubin (TSB). METHODS: Prospective blinded study comparing two diagnostic devices in 630 neonates requiring TSB measurement. RESULTS: Linear regression analysis showed a good correlation between BiliCheck™ and TSB (r=0.8212) as well as between JM-103 and TSB (r=0.8686). BiliCheck shows a tendency to underestimate TSB. The mean difference in TSB-TcB was -1.4 mg/dL for BC (-4.7/+1.8 mg/dL) and 0.3 mg/dL for JM-103 (-2.6/+3.2mg/dL). ROC analysis for TSB≥ 12 mg/dL showed area under the curve for BiliCheck™ significantly lower than those for JM-103 (p<0.0001). JM-103 resulted less time expensive than BiliCheck. CONCLUSIONS: In spite of similar diagnostic accuracy JM-103 could be preferred for some practical advantages, but its suitability in performing universal screening for severe hyperbilirubinemia deserves further investigations.
2012
Inglese
Romagnoli, C., Zecca, E., Catenazzi, P., Barone, G., Zuppa, A. A., Transcutaneous bilirubin measurement: Comparison of Respironics BiliCheck and JM-103 in a normal newborn population, <<CLINICAL BIOCHEMISTRY>>, 2012; 45 (9): 659-662. [doi:10.1016/j.clinbiochem.2012.03.014] [http://hdl.handle.net/10807/21374]
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/10807/21374
Citazioni
  • ???jsp.display-item.citation.pmc??? 11
  • Scopus 33
  • ???jsp.display-item.citation.isi??? 33
social impact